Carbolite Digital Furnace, Volatile Matter Furnace (ISO 562)

Carbolite VMF1006-volts

Prices Start From Call for price

Prices shown are for the United Arab Emirates only and are excluding VAT @ 5% Local Rate (Unless Otherwise Stated).

Other countries may incur additional import duties – Please Contact Us for further details and pricing for your location.

Link Gulf VMF10/6/3216P1 
Volatile Matter in Coal & Coke  ISO 562
Carbolite VMF1006-volts
 

  • This furnace offers improved temperature and reduced response times to meet the requirements of ISO 562;
  • Fast heating: Powerful open spiral elements are supported in lightweight insulation, providing fast heating – typically 20 minutes to 900°C;
  • Use of a sophisticated algorithm ensures fast recovery of temperature after loading the samples – less than 4 minutes to return to 900°C +10°C;
  • Calibration ports are provided which permit the insertion of unsheathed probe thermocouples, from the back of the chamber, without risking touching live electrical connections;
  • The probe thermocouple can be positioned under each crucible in turn;
  • A chimney or flue is provided at the back of the chamber;
  • Insulation: Hardwearing refractory brick is used around the doorway and a SiC floor provides abrasion resistance : lightweight ceramic fibre is used in other areas to ensure energy efficient fast heating and cooling, together with low heat loss into the room;
  • Door action: A parallel upward action keeps the hot face insulation away from the operator during loading and unloading : when the door is opened, a positive action safety switch automatically isolates the heating elements from the supply to protect the operator from accidental contact with a live heating element;
  • Power switching : solid state thyristor based zero voltage switching and rapid cycle time give smooth and reliable control with minimal electromagnetic emissions;
  • Service is aided by diagnostic warning lights, a removable front instrument panel and easy access to the heating elements and thermocouple;
  • Cool & safe outer case temperatures are achieved by natural convection moving air silently and reliably through cooling passages.



 

The VMF is specifically designed for testing the volatile matter of coal.

Standard features

  • VMF / ASTM for testing to ASTM D3175-11
  • VMF / ASTM complete with crucible, lid and holder
  • VMF 10/6 for testing to ISO 562:2010
  • VMF 10/6 parallel opening door
  • VMF 10/6 calibration ports to enable unsheathed thermocouple to be positioned below each crucible

Options (specify these at time of order)

  • Accessory crucibles and lids available
  • 4 crucible rack with loading handle (for VMF 10/6)
  • 9 crucible rack with loading hangle (for VMF 10/6)
  • Over temperature protection
  • Digital communications or nanodac paperless chart recorder to provide documentary evidence of the test procedure (for VMF 10/6)
  • Optional calibrated thermocouple probe with digital indicator (for VMF 10/6)

Content may be subject to modifications or corrections

VMF 10/6

Type Front loading
Maximum continuous operating temp (°C) 900
Heat-up time (mins) 20 to 900°C
Dimensions: Internal H x W x D (mm) 100 x 210 x 260
Dimensions:
External H x W x D (mm)
655 x 435 x 610
Max power (W) 3000
Thermocouple type K
Weight (kg) 47
Power supply 220 – 240V, 50-60Hz, single phase, 13A

  
Coal & Coke Testing Chamber Furnace

Catalog Numbers Description Phase Voltage

VMFASTM-volts
VMFASTM-120SN VMF/ASTM D3175 1000°C, ASTM D3175 single 110 – 120V
VMFASTM-230SN VMF/ASTM D3175 1000°C, ASTM D3175 single 220 – 240V
VMF1006-volts *
VMF1006-208SN VMF 10/6 /ISO562 1000°C, ISO562 single 200 – 210V
VMF1006-230SN VMF 10/6 /ISO562 1000°C, ISO562 single 220 -240V

  