Home » Entire Product Range » Refractometers » Hanna Instruments Salinity Refractometer Sodium Chloride Refractometer
Hanna Instruments HI96821
AVAILABLE IN STOCK
Prices Start From Call for price
Prices shown are for the United Arab Emirates only and are excluding VAT @ 5% Local Rate (Unless Otherwise Stated).
Other countries may incur additional import duties – Please Contact Us for further details and pricing for your location.
Description:
HANNA offers the HI96821 digital sodium chloride refractometer to meet the requirements of the food industry. This optical instrument that employs the measurement of the refractive index to determine sodium chloride concentration in aqueous solutions used in food preparation. It is not intended for sea water salinity measurements. The measurement of refractive index is simple and quick and provides the user an accepted method for NaCl analysis. Samples are measured after a simple user calibration with deionized or distilled water. Within seconds the instrument measures the refractive index of the solution. The digital refractometer eliminates the uncertainty associated with mechanical refractometers and is easily portable for measurements where you need them. The instrument utilizes internationally recognized references for unit conversion and temperature compensation. It can display the measurement of NaCl concentration 4 different ways: g/100 g, g/100 mL, Specific Gravity, and °Baumé. Temperature (in °C or °F) is displayed simultaneously with the measurement (on 3 of the ranges) on the large dual level display along with icons for Low Power and other helpful message codes.
|
Range
|
g/100 g
|
0 to 28
|
Range
|
g/100 mL
|
0 to 34
|
Range
|
Specific Gravity
|
1.000 to 1.216
|
Range
|
°Baumé
|
0 to 26
|
Range
|
Temperature
|
0 to 80°C (32 to 176°F)
|
Resolution
|
g/100 g
|
0.1
|
Resolution
|
g/100 mL
|
0.1
|
Resolution
|
Specific Gravity
|
0.001
|
Resolution
|
°Baumé
|
0.1
|
Resolution
|
Temperature
|
±0.1°C (0.1°F)
|
Accuracy (@20°C)
|
g/100 g
|
±0.2
|
Accuracy (@20°C)
|
g/100 mL
|
±0.2
|
Accuracy (@20°C)
|
Specific Gravity
|
±0.002
|
Accuracy (@20°C)
|
°Baumé
|
±0.2
|
Accuracy (@20°C)
|
Temperature
|
±0.3°C (0.5°F)
|
Temperature Compensation
|
|
Automatic between 10 and 40°C (50 to 104°F)
|
Measurement Time
|
|
Approximately 1.5 seconds
|
Minimum Sample Volume
|
|
100 µL (to cover prism totally)
|
Light Source
|
|
Yellow LED
|
Sample Cell
|
|
Stainless steel ring and flint glass prism
|
Auto-off
|
|
After 3 minutes of non-use
|
Enclosure Rating
|
|
IP 65
|
Battery Type / Battery Life
|
|
9V / Approx 5000 readings
|
Dimensions
|
|
192 x 102 x 67 mm (7.6 x 4 x 2.6”)
|
Weight
|
|
420 g
|