Home » Link Gulf – Main Store Page » Thermometers » IKA ETS-D5/D6 Electronic Contact Thermometer – IKA 3378100
– With optimized PID control
– RESET function
– With an integrated pH measuring instrument
– Incl. sensor H 62
– For all magnetic stirrers with contact thermometer bushing according to DIN 12878, class 2 (e.g. IKA®, Heidolph, and Corning with adapter AD-C1)
– With 3 operating modes
– Ensures perfect temperature control without overshooting the set temperature, even in the case of quick heating
– Large, graphic LCD with multilingual user guide
– Without pH electrode
Request a Quote
Prices shown are for the United Arab Emirates only and are excluding VAT @ 5% Local Rate (Unless Otherwise Stated).
Other countries may incur additional import duties – Please Contact Us for further details and pricing for your location.
– With optimized PID control
– RESET function
– With an integrated pH measuring instrument
– Incl. sensor H 62
– For all magnetic stirrers with contact thermometer bushing according to DIN 12878, class 2 (e.g. IKA®, Heidolph, and Corning with adapter AD-C1)
– With 3 operating modes
– Ensures perfect temperature control without overshooting the set temperature, even in the case of quick heating
– Large, graphic LCD with multilingual user guide
– Without pH electrode
3 modes of operation guarantee optimum adjustment to your working method:
Operating mode A: Suitable for work with varying parameters (from -50 °C to 450 °C). Safety temperature adjustable
Operating mode B: Suitable for series operation under uniform conditions
Operating mode C: Suitable for unsupervised operation
All values are taken from the memory. This ensures perfect protection against inadvertent improper adjustment
| Temperature measuring range min. | -50 °C |
| Temperature measuring range max. | 450 °C |
| Temperature measurement resolution | .01 K |
| Accuracy of temperature measurement | ±0.05 + tolerance PT1000 (DIN IEC 751 Class A) K |
| Heat control accuracy | .1 ±K |
| Control accuracy with sensor | .2 ±K |
| pH measuring range min. | 0 pH |
| pH measuring range max. | 14 pH |
| Accuracy of pH measurement | .1 pH |
| pH measurement resolution | .01 pH |
| Immersion depth max. | 200 mm |
| Permissible ON time | 100 % |
| Dimensions (W x H x D) | 96 x 98 x 45 mm |
| Weight | .2 kg |
| Permissible ambient temperature | 0 – 60 °C |
| Permissible relative moisture | 80 % |
| Protection class according to DIN EN 60529 | IP 54 |
| RS 232 interface | no |
| Analog output | no |
| DC Voltage | 8 – 16 V= |
| Power input | 15 mA |