Jenway 4510 Bench Conductivity Meter

Jenway 451 101 / Cole-Parmer EW-35440-10 / Link Gulf 451101

The Jenway-4510 Bench Conductivity Meter is easy to use, but with the flexibility to meet the broadest range of applications, and for those where greater accuracy is required, the 4510 has automatic conductivity standard recognition and endpoint detection. Set-up options include cell constant, temperature coefficient and reference temperature. The instrument can store up to 32 readings, which can also be sent to a printer or transferred to a computer via DataWay and the RS232 interface.

  • Simultaneous display of conductivity or TDS and temperature
  • Automatic selection of range for best resolution
  • Calibration by cell constant or standard solutions
  • Automatic standard recognition
  • Storage of up to 32 results
  • Bi-directional RS232 link to printer or PC
  • 3 year warranty

REQUEST A QUOTE

Prices shown are for the United Arab Emirates only and are excluding VAT @ 5% Local Rate (Unless Otherwise Stated).

Other countries may incur additional import duties – Please Contact Us for further details and pricing for your location.

Jenway-4510 Bench Conductivity Meter
Jenway 451 101 / Cole-Parmer EW-35440-10 / Link Gulf 451101

Jenway-4510

The Jenway-4510 Bench Conductivity Meter is easy to use, but with the flexibility to meet the broadest range of applications, and for those where greater accuracy is required, the 4510 has automatic conductivity standard recognition and endpoint detection. Set-up options include cell constant, temperature coefficient and reference temperature. The instrument can store up to 32 readings, which can also be sent to a printer or transferred to a computer via DataWay and the RS232 interface.

  • Simultaneous display of conductivity or TDS and temperature
  • Automatic selection of range for best resolution
  • Calibration by cell constant or standard solutions
  • Automatic standard recognition
  • Storage of up to 32 results
  • Bi-directional RS232 link to printer or PC
  • 3 year warranty

Jenway-4520

The 4520 is a high specification laboratory conductivity /resistivity/TDS/salinity and temperature meter that offers additional accuracy with 1, 2 or 3 point conductivity calibration across the wide measurement range of 0 to 19.99S. The dedicated “pure water” mode ensures the optimum accuracy for this difficult application.

  • Conductivity, resistivity, TDS and salinity modes
  • Special pure water mode for ultra low conductivity measurements
  • Storage of 500 results with built-in data logger
  • Infra-red data link (IrDA) for connection to printer and RS232 link to connect to printer or PC via DataWay software
  • 3 year warranty

 

Catalog Number Description Voltage
451101 4510 conductivity meter supplied with glass conductivity probe with ATC (K= 1) (027 013), electrode holder and US power supply 110V
452101 * 4520 conductivity meter supplied with glass conductivity probe with ATC (K= 1) (027 013), electrode holder and US power supply 110V
451201 4510 conductivity meter supplied with glass conductivity probe with ATC (K= 1) (027 013), electrode holder and EU power supply 230V EU Plug
452201 4520 conductivity meter supplied with glass conductivity probe with ATC (K= 1) (027 013), electrode holder and EU power supply 230V EU Plug
451001 4510 conductivity meter supplied with glass conductivity probe with ATC (K=1) (027 013), electrode holder and UK power supply 230V UK Plug
452001 4520 conductivity meter supplied with glass conductivity probe with ATC (K=1) (027 013), electrode holder and UK power supply 230V UK Plug

 

Parameter 4510 4520
Conductivity range 6 auto–selected from 0 to 1999mS* 6 auto–selected from 0 to 19.99S*
Conductivity resolution 0.01µS to 1mS 0.01µS to 0.01S
Conductivity accuracy ±0.5% ±2 digits ±0.5% ±2 digits
Automatic standard recognition 10µS, 84µS, 1413µS, 12.88mS 10µS, 84µS, 1413µS, 12.88mS
Cell constant 0.010 to 19.999 0.010 to 19.999
TDS range 6 auto-selected from 0 –1999g/l* 6 auto selected ranges from 0 –1999g/l*
TDS resolution 0.01mg/l to 1g/l 0.01mg/l to 1g/l
TDS accuracy ±0.5% ±2 digits ±0.5% ±2 digits
EC Ratio 0.50 to 0.80 0.50 to 0.80
Resistivity range NA 0–20Mohm
Resistivity resolution NA 0.01Mohm
Resistivity accuracy NA ±0.5% ±2 digits
Salinity range NA 0–99.9g/l
Salinity resolution NA 0.1g/l
Salinity accuracy NA 0–35 ±1g/l, 35–99.9 ±3g/l
Temperature range –10 to 105°C –10 to 105°C
Temperature resolution 0.1°C 0.1°C
Temperature accuracy ±0.5°C ±0.5°C
ATC and manual 0 to 100°C 0 to 100°C
Temperature coefficient 0.00% to 4.00%/°C 0.00% to 4.00%/°C
Reference temperature 18, 20 or 25°C 18, 20 or 25°C
GLP NA Calibration reminder interval, alarm outputs and security code protected data
Outputs Analogue and RS232 Analogue, RS232 and IrDA interface
Connector 7-pin DIN 7–pin DIN
Power 9V AC ±10% @ 50/60Hz 9V AC ±10% @ 50/60Hz
Size (l x w x d) 210 x 250 x 55mm 210 x 250 x 55mm
Weight 850g 850g