LabTech 9100 series UV-Vis spectrophotometers are designed for the diverse needs of environmental, pharmaceutical, life science, agricultural and clinical laboratories, as well as routine applications such as spectrum scanning, photometric measurement, kinetic, quantitative analysis, and multiple components analysis etc.
LabTech UV-Vis spectrophotometers offer fast scanning, excellent sensitivity, exceptional ruggedness and reliability, minimum maintenance, and excellent optical performance. Standard automatic 6-position cell holder, double beam optical system, five slit types, 0.5nm, 1nm, 2nm, 3nm, and 5nm, powerful and user friendly software provides packages for water quality test and biochemical test. Optional standalone or PC control mode for flexible operation.
Technical Features
Real double-beam system. High integration 32Bit CPU guarantees high performance. Five kinds of slits: 0.5nm, 1nm, 2nm, 3nm, and 5nm. Quartz coated and sealed optics prevents corrosion from environment. Light sources automatically record the working time. Automated six-coupled cell enormously shortens the sample replacement time. Intellectual auto-checking function of the software is convenient for trouble shooting and maintenance. Standalone systems are with powerful 32M storage capacity. Options of standalone systems or systems with separate PC controllers. Streamline surface protects sample splashes into internal space.
|
Model
|
UV9100A
|
UV9100B
|
UV9100C
|
UV9100D
|
Product Number
|
U9A-110
U9A-230
|
U9B-110
U9B-230
|
U9C-110
U9C-230
|
U9D-110
U9D-230 *
|
Bandwidth
|
2nm
|
1nm, 2nm
|
0.5nm, 2nm
|
0.5nm, 1nm, 2nm, 3nm, 5nm
|
Optical System
|
Double Beam, C-T monochromator, holographic grating
|
Wavelength Range
|
190 ~1100nm
|
Photometric Method
|
Transmittance, Absorbance, Energy
|
Wavelength Accuracy
|
±0.3nm
|
Wavelength Repeatability
|
±0.1nm
|
Absorbance Range
|
-3 ~ 3Abs
|
Photometric Accuracy
|
±0.002A (0 ~ 0.5A), ±0.004A (0.5 ~ 1A), ±0.3%T
|
Photometric Repeatability
|
≤0.001A (0 ~ 0.5A), ≤0.002A (0.5 ~ 1A), ±0.1%T
|
Wavelength Scan Speed
|
Adjustable (Fast, Middle, Low)
|
Stray Light
|
≤0.05% T
|
Baseline Straightness
|
±0.001A
|
Drift
|
≤0.0003A/h (500nm, 0A)
|
Noise
|
±0.0002A (500nm, P-P)
|
Detector
|
Si-Photodiode
|
Light Source
|
Halogen tungsten lamp / Deuterium lamp (Auto-switch)
|
Cell Holder
|
Automatic six-position cell holder
|
Display
|
320 × 240 pixel, LCD
|
Control Mode
|
Local control or PC control
|
Dimensions (WxDxH) Inch
|
20x19x10
|
Weight Lbs
|
49
|
|
Model
|
UV9100A
|
UV9100B
|
UV9100C
|
UV9100D
|
Product Number
|
U9A
|
U9B
|
U9C
|
U9D *
|
Bandwidth
|
2nm
|
1nm, 2nm
|
0.5nm, 2nm
|
0.5nm, 1nm, 2nm, 3nm, 5nm
|
Optical System
|
Double Beam, C-T monochromator, holographic grating
|
Wavelength Range
|
190 ~1100nm
|
Photometric Method
|
Transmittance, Absorbance, Energy
|
Wavelength Accuracy
|
±0.3nm
|
Wavelength Repeatability
|
±0.1nm
|
Absorbance Range
|
-3 ~ 3Abs
|
Photometric Accuracy
|
±0.002A (0 ~ 0.5A), ±0.004A (0.5 ~ 1A), ±0.3%T
|
Photometric Repeatability
|
≤0.001A (0 ~ 0.5A), ≤0.002A (0.5 ~ 1A), ±0.1%T
|
Wavelength Scan Speed
|
Adjustable (Fast, Middle, Low)
|
Stray Light
|
≤0.05% T
|
Baseline Straightness
|
±0.001A
|
Drift
|
≤0.0003A/h (500nm, 0A)
|
Noise
|
±0.0002A (500nm, P-P)
|
Detector
|
Si-Photodiode
|
Light Source
|
Halogen tungsten lamp / Deuterium lamp (Auto-switch)
|
Cell Holder
|
Automatic six-position cell holder
|
Display
|
320 × 240 pixel, LCD
|
Control Mode
|
Local control or PC control
|
Dimensions (WxDxH) Inch
|
20x19x10
|
Weight Lbs
|
49
|
|